Combined active and passive breathing
simulator with digitally controlled piston
for simulation of compliances and
resistances via a pressure feedback loop.
(Part No. 31 00 100) |
|
Modes of Operation
|
Parameter file scripts
Interactive (parameters changed directly by
the user)
Time-varying parameters
(parameter changes adjusted breath by breath
in a pre-scripted fashion)
External interactive
(parameter change algorithms provided by
user-written software)
Pause
mode (user-definable idling state)
|
|
Lung Model Types |
Passive (1 or 2 compartment)
Spontaneous
Combined (with adjustable response to
ventilator breaths)
Flow waveform generator
|
|
Hardware
Total volume |
3.1 L |
|
Tidal volumes |
2 to 2700 cm3
“FRC” setting adjustable (200 –
1500 cm3) |
|
Spontaneous
Breath rate |
0 to 150 / min (infant Vt) |
|
Peak flow |
280 L / min ± 10% (t90Flow < 50 ms) |
|
Resistance settings |
3 to 500 cm H2O
/ L / s, (linear and parabolic resistor
types) |
|
Compliance settings |
0.5 to 1000 mL /
cm H2O |
|
Breath profiles |
user-defined pressure data files (breath
profiles)
or parameter-selected standard patterns |
|
Leak settings |
possible through optional SBLVM Simulator Bypass and
Leak Valve Module |
|
Small signal |
better than 15 Hz bandwidth
(10 cm3
response to HF ventilation) |
|
Airway pressure |
uncertainty < than 1% fso measurement |
|
Barometric pressure |
uncertainty < 1% (1 kPa) |
|
Gas temperature |
uncertainty < 0.5 Co (20 – 45 Co) |
|
Input/Output
Input |
2 channels analog input 0-10 V 16 bit
solution
(can be used for auxiliary pressure and flow
signals) |
|
Output |
5 V TTL-signal
for breath phase recognition, (may be used
for synchronizing external data acquisition
system with the ASL 5000 simulator) |
|
Software
Interface to PC Host |
via RS 232 at 57.6 kBaud |
|
Host |
LabVIEWTM
based software environment
With an integrated set of applications for:
-modeling-simulation
-data analysis |
|
Servo update rate |
2048 Hz |
|
Electrical |
Universal input, 90…240 V AC |
|
Dimensions |
425 x 219 x 315 mm |
|
Weight |
10 kg (approx.) 22 lbs. |
|
Patent |
Protected under US patent 5975748 |
|
 |
|
|
Simulator Bypass and Leak Valve Module (SBLVM)
External accessory for setting 3 levels of
leaks. Also switches between simulator and
an external breathing bag when no simulation
is running to avoid
alarming ventilator.
(Part No. 31 00 200) |
|
Leak settings |
approximately 4, 9, 15 L/min at
20 cm H2O, leak orifices
exchangeable |
|
Dimensions |
4.8 x 4.8 x 4.3 inches / 123 x 123 x 108 mm |
|
Weight |
3.5 lbs. (1.6 kg) |
|
Fast Oxygen Measurement (FOM)
For comprehensive ventilator evaluation.
Delivers reading almost instantaneously from
a highly accurate paramagnetic oxygen
sensor in sidestream configuration. Integrated sample
pump and barometric pressure sensor for
pressure correction.
(Part No. 31 00 300) |
|
Measurement range |
0 to 100% |
|
Uncertainty |
less than
+ 0.5% O2 |
|
Response time |
<350 msec (t10/90
21 to 100% O2) |
|
Cylinder Temperature Controller (CTC)
For simulations requiring realistic
physiological conditions. Foil heater with
auto-tune PID controller, wall temperature
probe, and independent safety thermostat
(120 or 240 V).
(Part No. 31 00 400) |
|
Wall temperature
Setting |
ambient to 45o
C |
|
Carrying Case
Hardshell transport case with foam insert,
retractable handle and wheels.
(Part No. 31 00 500)
|
|
Dimensions |
25 x 20 x 14 inches / (63 x 51 x 36 cm) |
|
Color: |
black |
|
Auxiliary Gas Exchange Cylinder (AGEC)
Bag-in-bottle accessory for use of ASL 5000
with aggressive chemicals.
Designed to work with bags and bellows < 5”
OD.
(Part No. 31 00 600) |
|
Volume |
approximately 3L |
|
Mobile Cart
Ergonomic design holds laptop or a digital
projector and 17 or 18" LCD flat panel
display. Four twin wheel swivel castors for
maximum stability.
(Part No. 31 00 700)
|